Hanna Instruments Boiler and Cooling Tower Photometer – HI83305
The HI83305 is a small multiparameter photometer designed for use in the lab or in the field for boiling and cooling tower applications. The meter is one of the most advanced available, with an innovative optical design that employs a reference detector and focusing lens to eliminate errors caused by changes in the light source as well as imperfections in the glass cuvette. This meter includes 30 different programmed methods for measuring 18 key water quality parameters, as well as an absorbance measurement mode for performance verification and users who want to develop their own concentration versus absorbance curves. Specific parameters for boiling and cooling towers include oxygen scavengers and silica, which are important for equipment maintenance.
With its digital pH/temperature electrode input, the HI83305 doubles as a professional pH meter, saving valuable laboratory benchtop space. Now, a single meter can perform both photometric and pH measurements.
- A cutting-edge optical system
- A benchtop photometer with unrivaled performance
- Input from a digital pH electrode
- Save valuable bench space by using a meter that functions as both a photometer and a laboratory pH meter.
SKU | HI83305-02 |
---|---|
Product Name | Boiler and Cooling Tower Photometer – HI83305 |
Quote Required | Yes |
pH Range | Photometer |
pH Resolution | Photometer |
pH Accuracy | Photometer |
pH Calibration | Automatic one or two point calibration with one set of standard buffers available (4.01, 6.86, 7.01, 9.18, 10.01) |
pH Temperature Compensation | Automatic (-5.0 to 100.0 oC; 23.0 to 212.0 oF); limits reduced based on the pH electrode used |
pH CAL Check (electrode diagnostics) | clean electrode and check buffer/check probe displayed during calibration |
pH Method | Photometer |
pH-mV Range | ±1000 mV |
pH-mV Resolution | 0.1 mV |
pH-mV Accuracy | ±0.2 mV |
Dissolved Oxygen Range | 0.0 to 10.0 mg/L (as O2) |
Dissolved Oxygen Resolution | 0.1 mg/L |
Dissolved Oxygen Accuracy | ±0.4 mg/L ±3% of reading |
Dissolved Oxygen Measurement Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Azide modified Winkler method |
Absorbance Range | 0.000 to 4.000 Abs |
Absorbance Resolution | 0.001 Abs |
Absorbance Accuracy | +/-0.003Abs @ 1.000 Abs |
Aluminum Range | 0.00 to 1.00 mg/L (as Al3+) |
Aluminum Resolution | 0.01 mg/L |
Aluminum Accuracy | ±0.04 mg/L ±4% of reading |
Aluminum Method | Adaptation of the aluminon method |
Ammonia Range | Low Range |
Ammonia Resolution | 0.01 mg/L; 0.1 mg/L |
Ammonia Accuracy | Low Range |
Ammonia Method | Adaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method |
Bromine Range | 0.00 to 8.00 mg/L (as Br2) |
Bromine Resolution | 0.01 mg/L |
Bromine Accuracy | ±0.08 mg/L ±3% of reading |
Bromine Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method. |
Chlorine Dioxide Range | 0.00 to 2.00 mg/L (as ClO2) |
Chlorine Dioxide Resolution | 0.01 mg/L |
Chlorine Dioxide Accuracy | ±0.10 mg/L ±5% of reading |
Chlorine Dioxide Method | Adaptation of the Chlorophenol Red method. |
Free Chlorine Range | 0.00 to 5.00 mg/L (as Cl2) |
Free Chlorine Resolution | 0.01 mg/L |
Free Chlorine Accuracy | ±0.03 mg/L ±3% of reading |
Total Chlorine Range | 0.00 to 5.00 mg/L (as Cl2) |
Total Chlorine Resolution | 0.01 mg/L |
Total Chlorine Accuracy | ±0.03 mg/L ±3% of reading |
Chlorine Method | Adaptation of the EPA 330.5 DPD method |
Chromium, Hexavalent Range | Low Range |
Chromium, Hexavalent Resolution | 1 μg/L |
Chromium, Hexavalent Accuracy | Low Range |
Chromium, Hexavalent Method | Adaptation of the ASTM Manual of Water and Environmental Technology, D1687-92, Diphenylcarbohydrazide method. |
Copper Range | Low Range |
Copper Resolution | 0.001 mg/L; 0.01 mg/L |
Copper Accuracy | Low Range |
Copper Method | Adaptation of the EPA bicinchoninate method |
Hydrazine Range | 0 to 400 μg/L (as N2H4) |
Hydrazine Resolution | 1 μg/L |
Hydrazine Accuracy | ±4% of full scale reading |
Hydrazine Method | Adaptation of the ASTM Manual of Water and Environmental Technology, method D1385-88, p-Dimethylaminobenzaldehyde method |
Iron Range | Low Range |
Iron Resolution | 0.001 mg/L; 0.01 mg/L |
Iron Accuracy | Low Range |
Iron Method | Low Range |
Molybdenum Range | 0.0 to 40.0 mg/L (as Mo6+) |
Molybdenum Resolution | 0.1 mg/L |
Molybdenum Accuracy | ±0.3 mg/L ±5% of reading |
Molybdenum Method | Adaptation of the mercaptoacetic acid method |
Nitrate Range | 0.0 to 30.0 mg/L (as NO3–– N) |
Nitrate Resolution | 0.1 mg/L |
Nitrate Accuracy | ±0.5 mg/L ±10% of reading |
Nitrate Method | Adaptation of the cadmium reduction method |
Nitrite Range | Freshwater Low Range |
Nitrite Resolution | Freshwater |
Nitrite Accuracy | Freshwater Low Range |
Nitrite Method | Low Range |
Oxygen, Scavenger Range | 0 to 1000 μg/L (as DEHA) 0.00 to 1.50 mg/L (as Carbohydrazide) 0.00 to 2.50 mg/L (as Hydroquinone) 0.00 to 4.50 mg/L (as ISO-ascorbic acid) |
Oxygen, Scavenger Resolution | 1 μg/L (DEHA); 0.01 mg/L |
Oxygen, Scavenger Accuracy | ±5 μg/L ±5% of reading |
Oxygen, Scavenger Method | Adaptation of the iron reduction method |
Phosphate Range | Freshwater Low Range |
Phosphate Resolution | Freshwater |
Phosphate Accuracy | Freshwater Low Range |
Phosphate Method | Freshwater Low Range |
Silica Range | Low Range |
Silica Resolution | 0.01 mg/L; 1 mg/L |
Silica Accuracy | Low Range |
Silica Method | Low Range |
Zinc Range | 0.00 to 3.00 mg/L (as Zn) |
Zinc Resolution | 0.01 mg/L |
Zinc Accuracy | ±0.03 mg/L ±3% of reading |
Zinc Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Zincon method |
Input Channels | 1 pH electrode input and 5 photometer wavelengths |
pH Electrode | digital pH electrode (not included) |
Photometer/Colorimeter Light Source | 5 LEDs with 420 nm, 466 nm, 525 nm, 575 nm, and 610 nm narrow band interference filters |
Photometer/Colorimeter Light Detector | silicon photodetector |
Bandpass Filter Bandwidth | 8 nm |
Bandpass Filter Wavelength Accuracy | ±1 nm |
Cuvette Type | round, 24.6 mm |
Number of Methods | 128 max. |
GLP | calibration data for connected pH electrode |
Display | 128 x 64 pixel LCD with backlight |
Logging Type | log on demand with user name and sample ID optional input |
Logging Memory | 1000 readings |
Connectivity | USB-A host for flash drive; micro-USB-B for power and computer connectivity |
Power Supply | 5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included) |
Battery Type/Life | 3.7 VDC Li-polymer rechargeable battery / >500 photometric measurements or 50 hours of continuous pH measurement |
Environment | 0 to 50.0 oC (32 to 122.0 oF); 0 to 95% RH, non-condensing |
Weight | 1.0 kg (2.2 lbs.) |
Dimensions | 206 x 177 x 97 mm (8.1 x 7.0 x 3.8″) |
Ordering Information | HI83305 is supplied with sample cuvettes and caps (4 ea.), cloth for wiping cuvettes, USB to micro USB cable connector, power adapter and instruction manual. |