E-Sampler Dual Ambient Monitor/Sampler
Features:
- Auto-Zero and Auto Span are programmable.
- Automatic (1–65 000 m/m3).
- Protocol for automatic flow control.
- Inside Battery (30 Hours Operation without heater & 10 Hours with heater.).
- Precision optical engine with laser diodes.
- 47mm integral analysis filter
- Temperature and atmospheric pressure.
- A data logger inside.
- Monitoring for PM10, PM2.5, PM1, and TSP.
- Weatherproof Aluminum Enclosure.
- Optics with purge-air protection.
- Completely independent.
- Filter replacement without tools.
Categories: Climatic/Environment Inspection, Particle Counter
Description
Specifications:
PARAMETER | SPECIFICATION |
Measurement Principles: | Particulate concentration by forward light scatter laser Nephelometer. Integrated 47mm low flow gravimetric filter sampler. |
Available Cut Points: | TSP Inlet Standard. PM10, PM2.5, and PM1 sharp-cut cyclone inlets available. |
Measurement Range: | 0 to 65 mg/m3 (0 to 65,530 mg/m3) dynamic range. 16 bit digital range. |
Nephelometer Accuracy: | ± 10% to gravimetric method typical when K-factored to local particulate type. |
Gravimetric Accuracy: | ± 8% of NIOSH 0600. |
Precision: | Greater of 3 µg/m3 or 2%. |
Data Storage Resolution: | 1 µg/m3 |
Data Storage Intervals: | User-Selectable 1, 5, 10, 15, 30, or 60 minute averages. |
Nephelometer Interval: | 1-second measurements, available on analog output and display. |
Sample Cycles: | Continuous operation or programmable scheduled sample runs. |
Particle Size Sensitivity: | 0.1 to 100 micron. Optimal sensitivity 0.5 to 10 micron particles. |
Laser Type: | Diode Laser, 5 mW, 670nm. Visible red. |
Long Term Stability: | 5% with clean optics. |
Flow Rate: | 2.0 liters/minute ± 0.1 lpm. Actual volumetric flow. |
Pump Type: | 10,000 hour brushless diaphragm sample pump and secondary purge pump. |
Gravimetric Filter Type: | 47mm disc filters (not included). Accepts standard FRM filter holder cartridges. |
Automatic Zero and Span: | User-selectable 15 min, 1 hour, 2 hour, 12 hour, or 24 hour intervals. 2.8 min cycle. |
Internal Battery: | Optional lead acid battery, 12V, 12 Amp-Hour. Yuasa NP12-12 or equivalent. |
Internal Battery Run Time: | Up to 30 hours with inlet heater off. Up to 10 hours with inlet heater on. |
Power Supply: | Universal 100-240 VAC input, 15 VDC output power supply included. Compatible with solar power kits or external batteries using optional DC power cable. |
Power Consumption: | 1.1 amps @ 12 VDC (15 Watts) max continuous draw, running with inlet heater on.
0.35 amps (4.2 Watts) running with inlet heater off. |
Operating Temperature: | 0 to +50°C . (Ambient Temperature Sensor Range -30 to +50°C). |
Barometric Pressure: | 60,000 to 104,000 Pascal pressure sensor range. |
Ambient Humidity Range: | 0 to 90% RH, non-condensing. |
Humidity Control: | Automatic 10 Watt inlet heater module controlled to sample RH setpoint. Sample RH sensor standard. Optional EX-593 ambient RH sensor available. |
Approvals: | CE, ISO-9001. Designed to agree with EPA Class I and Class III FRM/FEM particulate samplers and monitors. Not an EPA-designated equivalent method. |
User Interface: | Menu-driven interface with 4×20 character LCD display and dynamic keypad. |
Analog Voltage Output: | 0-1, 0-2.5, or 0-5 volt DC output. User-set range with 1-second real-time output. |
Serial Interface: | RS-232 2-way serial port for PC, datalogger, or modem communications. |
Alarm Contact Closure: | Normally closed contact closure relay output. Contact rating 0.5A @ 100V DC max. |
Compatible Software: | Comet™ (included), Air Plus™, terminal programs such as HyperTerminal® |
Alarm Reporting: | Available through serial port data files, display, and relay output. |
Memory: | 4369 data logger records (182 days @ 1 record/hr, 3 days @ 1 record/min). |
Factory Service Interval: | 24 Months typical, under continuous use in normal ambient air. |
Mounting Options: | Pole or wall mount bracket standard. Optional EX-905 tripod recommended. |
Unit Weight: | 6.4 kg (14 lbs) without tripod, battery, or optional accessories. |
Unit Dimensions: | 65cm high, 27cm wide, 16.5cm deep. (25.5” x 10.5” x 6.5”). With inlet assembly |
Related products
Climatic/Environment Inspection
Climatic/Environment Inspection
Climatic/Environment Inspection
Air Flows Tester
Climatic/Environment Inspection
Climatic/Environment Inspection
Air Flows Tester
Climatic/Environment Inspection