TES PROVA-123 Process Calibrator
- 4-20mA, 0-20mA, 0-24mA with 1 μA resolution
- 0-100mV with 10 μV resolution 0-12V with 1mV resolution
- Thermocouple K,J,E,T TYPE 1 o C 1 o F resolution
- Frequency 2 – 50,000Hz
Electrical Specifications: (23℃±5℃, 10 minutes after power is on)
mA DC Current: (1KΩMax. Load, 24V Loop Supply)
Range |
Resolution | Accuracy |
4 – 20mA, 0 – 20mA, 0 -24 mA | 1μA | ± 0.025% ±3μA |
Beeper warning when output is open and specified output > 1mA
mV, V DC Voltage:(1mA Supply Current)
Range |
Resolution | Accuracy |
0 – 100.00mV | 10μV | ± 0.05% ± 30μV |
0 – 10.000V | 1mV | ± 0.05% ± 3mV |
0 – 1.0000V | 100μV | ± 0.05% ± 300μV |
Beeper warning when output is short and specified voltage output > 10mV
K, J, E, T type Thermocouples:(1°C, 1°F Resolution, 1KΩ Load Min.)
Range | Accuracy | Range | Accuracy |
K: -200 to -100 °C | ± 2.0°C | K: -328 to -148°F | ± 3.6°F |
K: -100 to 0 °C | ± 1.1°C | K: -148 to 32°F | ± 1.8°F |
K: 0 to 1370 °C | ± 0.8°C | K: 32 to 2400°F | ± 1.5°F |
J: -100 to 0 °C | ± 0.9°C | J: -148 to 32°F | ± 1.6°F |
J: 0 to 760 °C | ± 0.7°C | J: 32 to 1400°F | ± 1.2°F |
E: -100 to 0 °C | ± 0.9°C | E: -148 to 32°F | ± 1.6°F |
E: 0 to 700 °C | ± 0.7°C | E: 32 to 1292°F | ± 1.2°F |
T: -200 to 0 °C | ± 1.0°C | T: -328 to 32°F | ± 1.8°F |
T: 0 to 400 °C | ± 0.8°C | T: 32 to 752°F | ± 1.5°F |
Frequency:(1 – 125Hz, TTL 0 – 5V, 1KΩ Load Min.)
Range | Resolution | Accuracy |
1 – 125 Hz | 1 Hz | ± 0.04Hz |
Available Frequencies
(126 – 62500Hz, TTL 0 – 5V, Accuracy ± 0.01% ± 0.04Hz, 1KΩ Load Min):
126.00 | 127.03 | 128.07 | 129.13 | 130.20 | 131.30 | 132.41 | 133.54 | 134.12 | 135.28 |
136.46 | 137.06 | 138.27 | 139.50 | 140.13 | 141.40 | 142.04 | 143.34 | 144.00 | 145.34 |
146.02 | 147.40 | 148.10 | 149.52 | 150.24 | 151.69 | 152.43 | 153.18 | 154.70 | 155.47 |
………………………………………….
16666.66 17241.37 17857.14 18518.51 19230.76 20000.00 20833.33 21739.13 22727.27
23809.52 25000.00 26315.78 27777.77 29411.76 31250.00 33333.33 35714.28 38461.53
41666.66 45454.54 50000.00 55555.55 62500.00
General Specifications:
Dimension: | 88 x 168 x 26 mm
(3.46″ x 6.61″ x 1.03″) |
Battery Type: | 9V Alkaline Battery |
Display: | 4 and 5 Digits |
Power Consumption: | 60mA – 180mA |
Operating Temperature: | 0C to 50C
(32°F to 122°F) |
Operating Humidity: | less than 85% relative |
Storage Temperature: | -20C to 60C
(-4°F to 140°F) |
Storage Humidity: | less than 85% relative |
Weight: | 330g / 11.63 oz |
Accessories: | Carrying case x 1
Users Manual x 1 External Battery Pack: 1.5V AA Batteries x 6 9V Alkaline Battery x 1 Alligator clip x 1 K type thermocouple connector Test leads x 1 |