TES Datalogger Multimeter TES-2732
Datalogger Multimeter
- Easy Data Acquisition (Field/Laboratory)
- DMM talks with P.C via RS-232 Interface
- Data Logger function for Field recording without P.C can record up to 4048 data
- Drect input of V,A, Ohm, Hz, Cx & °C/°F without any further signal modification
- Go/No Go (alarm output) used for control purpose
Electrical Specifications
Accuracies are ±(…% of reading + …digits) at 23°C±5°C, below 80% RH
DC Voltage
Range | Resolution | Accuracy | Input Impedance | Overload Protection |
200mV | 0.1mV |
0.5%+1 |
10MΩ |
600Vrms |
2V | 1mV |
1100Vrms |
||
20V | 10mV | |||
200V | 100mV | |||
1100V | 1V | 0.5%+2 |
AC Voltage (50Hz to 500Hz)
Range | Resolution | Accuracy | Input Impedance | Overload Protection |
200mV | 0.1mV |
1%+5 |
10MΩ |
600Vrms |
2V | 1mV |
750Vrms |
||
20V | 10mV | |||
200V | 100mV | |||
750V | 1V | 1.2%+5 |
DC Current
Range | Resolution | Accuracy | Burden Voltage | Overload Protection |
200uA | 0.1uA |
1%+1 |
0.35V
|
0.5A/250V Fast Blow Fuse & Diode |
20mA | 1uA | |||
20mA | 10uA | |||
200mA | 100uA | |||
20A | 10mA | 1.2%+3 | 0.8V | 20A/380V Fast Blow Fuse
20A for 30sec max. 10A Continuity |
AC Current (50Hz to 500Hz)
Range | Resolution | Accuracy | Burden Voltage | Overload Protection |
200uA | 0.1uA |
1.2%+3 |
0.35V |
0.5A/250V Fast Blow Fuse & Diode |
2mA | 1uA | |||
20mA | 10uA | |||
200mA | 100uA | |||
20A | 10mA | 1.5%+5 | 0.8V | 20A/380V Fast Blow Fuse
20A for 30sec max. 10A Continuity |
Resistance (Ω)
Range | Resolution | Accuracy | Max. Open Circuit Voltage | Overload Protection |
200 Ω | 0.1 Ω |
0.8%+2 |
3.2V |
600Vrms |
2KΩ | 1 Ω |
0.5V |
||
20KΩ | 10 Ω | |||
200KΩ | 100 Ω | |||
2MΩ | 1KΩ | |||
20MΩ | 10KΩ | 1.5%+3 |
Frequency (Hz)(Test Range 10Hz – 10MHz Auto Range)
Range | Resolution | Accuracy | Sensitivity | Overload Protection |
2KHz | 1Hz |
0.5%+2 |
1.5Vrms |
600Vrms |
20KHz | 10Hz | |||
200KHz | 100Hz | |||
2MHz | 1KHz | |||
10MHz | 10KHz | 3Vrms |
Temperature Probe Used: K(CA) type sensor
Range | Resolution | Accuracy |
°C | 0.1/1°C | 0°C ~ 200°C ±(0.5%+1.5°C)
-50°C ~ 0°C ±(1%+2°C) 200°C ~ 1300°C ±(1%+2°C) |
°F | 0.1/1°F | 32°F ~ 200°F ±(0.5%+3°F)
-58°F ~ 32°F ±(1%+5°F) 200°F ~ 1999°F ±(1%+4°F) |
Capacitance (F)
Range | Resolution | Accuracy | Test Frequency |
2000pF | 1pF |
3%+10 |
400Hz |
200nF | 100pF | ||
2uF | 1nF | ||
20uF | 10nF |
Display:
3 ½ Digital LCD Display Maximum Reading 1999
Sampling Rate: 2.5 readings per second
Operating/Storage Condition:
0°C to 40°C<80% R.H.
-10°C to 60°C<70% R.H.
Power Source: One 9V Battery
Battery Life Approx. 60 hours (Alkaline Battery)
Size & Weight: 180(L)x82(W)x38(H)mm & 365g
Accessories: Test leads (pair), Spare fuse (0.5A/250V), Battery, Instruction Manual, Software, RS-232 Cable, “K” type thermocouple (TP-K01)